Misplaced Pages

Atomic force microscopy

Article snapshot taken from Wikipedia with creative commons attribution-sharealike license. Give it a read and then ask your questions in the chat. We can research this topic together.

Atomic force microscopy ( AFM ) or scanning force microscopy ( SFM ) is a very-high-resolution type of scanning probe microscopy (SPM), with demonstrated resolution on the order of fractions of a nanometer, more than 1000 times better than the optical diffraction limit .

#245754

101-414: Atomic force microscopy (AFM) is a type of SPM, with demonstrated resolution on the order of fractions of a nanometer, more than 1000 times better than the optical diffraction limit . The information is gathered by "feeling" or "touching" the surface with a mechanical probe. Piezoelectric elements that facilitate tiny but accurate and precise movements on (electronic) command enable precise scanning. Despite

202-631: A NA of 1, the Abbe limit is roughly d = λ 2 = 250  nm {\displaystyle d={\frac {\lambda }{2}}=250{\text{ nm}}} (0.25 μm), which is small compared to most biological cells (1 μm to 100 μm), but large compared to viruses (100 nm), proteins (10 nm) and less complex molecules (1 nm). To increase the resolution, shorter wavelengths can be used such as UV and X-ray microscopes. These techniques offer better resolution but are expensive, suffer from lack of contrast in biological samples and may damage

303-406: A ceramic material) (3) oscillates the cantilever (1). The sharp tip (4) is fixed to the free end of the cantilever (1). The detector (5) records the deflection and motion of the cantilever (1). The sample (6) is mounted on the sample stage (8). An xyz drive (7) permits to displace the sample (6) and the sample stage (8) in x, y, and z directions with respect to the tip apex (4). Although Fig. 3 shows

404-477: A combination of incoherent or structured illumination, as well as collecting both forward, and backward scattered light it is possible to image the complete scattering sphere . Unlike methods relying on localization , such systems are still limited by the diffraction limit of the illumination (condenser) and collection optics (objective), although in practice they can provide substantial resolution improvements compared to conventional methods. The diffraction limit

505-420: A composite of images illuminated from each point on the condenser, each of which covers a different portion of the object's spatial frequencies. This effectively improves the resolution by, at most, a factor of two. Simultaneously illuminating from all angles (fully open condenser) drives down interferometric contrast. In conventional microscopes, the maximum resolution (fully open condenser, at N = 1)

606-421: A constant amplitude of the cantilever oscillation as long as there is no drift or interaction with the surface. The interaction of forces acting on the cantilever when the tip comes close to the surface, van der Waals forces , dipole–dipole interactions , electrostatic forces , etc. cause the amplitude of the cantilever's oscillation to change (usually decrease) as the tip gets closer to the sample. This amplitude

707-451: A constant probe-sample interaction (see § Topographic image for more). The surface topography is commonly displayed as a pseudocolor plot. Although the initial publication about atomic force microscopy by Binnig, Quate and Gerber in 1986 speculated about the possibility of achieving atomic resolution, profound experimental challenges needed to be overcome before atomic resolution of defects and step edges in ambient (liquid) conditions

808-628: A conventional diffraction-limited objective, improving the axial resolution. However, because these techniques cannot image beyond 1 wavelength, they cannot be used to image into objects thicker than 1 wavelength which limits their applicability. Far-field imaging techniques are most desirable for imaging objects that are large compared to the illumination wavelength but that contain fine structure. This includes nearly all biological applications in which cells span multiple wavelengths but contain structure down to molecular scales. In recent years several techniques have shown that sub-diffraction limited imaging

909-424: A few picometers. The van der Waals forces , which are strongest from 1 nm to 10 nm above the surface, or any other long-range force that extends above the surface acts to decrease the resonance frequency of the cantilever. This decrease in resonant frequency combined with the feedback loop system maintains a constant oscillation amplitude or frequency by adjusting the average tip-to-sample distance. Measuring

1010-416: A given motor power). Potentiometers are subject to drift when the temperature changes whereas encoders are more stable and accurate. Servomotors are used for both high-end and low-end applications. On the high end are precision industrial components that use a rotary encoder. On the low end are inexpensive radio control servos (RC servos) used in radio-controlled models which use a free-running motor and

1111-404: A hardness of cells, and to evaluate interactions between a specific cell and its neighboring cells in a competitive culture system. AFM can also be used to indent cells, to study how they regulate the stiffness or shape of the cell membrane or wall. In some variations, electric potentials can also be scanned using conducting cantilevers. In more advanced versions, currents can be passed through

SECTION 10

#1732858475246

1212-407: A material stuck on the tip of the cantilever, and from another hand the surface of particles either free or occupied by the same material. From the adhesion force distribution curve, a mean value of the forces has been derived. It allowed to make a cartography of the surface of the particles, covered or not by the material. AFM has also been used for mechanically unfolding proteins. In such experiments,

1313-427: A microcontroller) sends pulse-width modulation (PWM) signals to the servo. The electronics inside the servo translate the width of the pulse into a position. When the servo is commanded to rotate, the motor is powered until the potentiometer reaches the value corresponding to the commanded position. James Watt 's steam engine governor is generally considered the first powered feedback system. The windmill fantail

1414-406: A quantitative manner from phase images, however, is often not feasible. In non-contact atomic force microscopy mode, the tip of the cantilever does not contact the sample surface. The cantilever is instead oscillated at either its resonant frequency (frequency modulation) or just above (amplitude modulation) where the amplitude of oscillation is typically a few nanometers (<10 nm) down to

1515-420: A resolution better than the diffraction limit by locating the objective lens extremely close (typically hundreds of nanometers) to the object. In fluorescence microscopy the excitation and emission are typically on different wavelengths. In total internal reflection fluorescence microscopy a thin portion of the sample located immediately on the cover glass is excited with an evanescent field, and recorded with

1616-828: A rotary (angular) or linear output. Speed control via a governor is another type of servomechanism. The steam engine uses mechanical governors; another early application was to govern the speed of water wheels . Prior to World War II the constant speed propeller was developed to control engine speed for maneuvering aircraft. Fuel controls for gas turbine engines employ either hydromechanical or electronic governing. Positioning servomechanisms were first used in military fire-control and marine navigation equipment. Today servomechanisms are used in automatic machine tools , satellite-tracking antennas, remote control airplanes, automatic navigation systems on boats and planes, and antiaircraft -gun control systems. Other examples are fly-by-wire systems in aircraft which use servos to actuate

1717-455: A simple potentiometer position sensor with an embedded controller. The term servomotor generally refers to a high-end industrial component while the term servo is most often used to describe the inexpensive devices that employ a potentiometer. Stepper motors are not considered to be servomotors, although they too are used to construct larger servomechanisms. Stepper motors have inherent angular positioning, owing to their construction, and this

1818-632: A single atom) has also been achieved by AFM. In manipulation, the forces between tip and sample can also be used to change the properties of the sample in a controlled way. Examples of this include atomic manipulation, scanning probe lithography and local stimulation of cells. Simultaneous with the acquisition of topographical images, other properties of the sample can be measured locally and displayed as an image, often with similarly high resolution. Examples of such properties are mechanical properties like stiffness or adhesion strength and electrical properties such as conductivity or surface potential. In fact,

1919-414: A single image with data covering a larger portion of the object's spatial frequencies when compared to using a fully closed condenser (which is also rarely used). Another technique, 4Pi microscopy , uses two opposing objectives to double the effective numerical aperture, effectively halving the diffraction limit, by collecting the forward and backward scattered light. When imaging a transparent sample, with

2020-537: A spatial or angular extent essentially equal to the resolution of the optics at the wavelength of the laser. The observation of sub-wavelength structures with microscopes is difficult because of the Abbe diffraction limit . Ernst Abbe found in 1873, and expressed as a formula in 1882, that light with wavelength λ {\displaystyle \lambda } , traveling in a medium with refractive index n {\displaystyle n} and converging to

2121-580: A spot with half-angle θ {\displaystyle \theta } will have a minimum resolvable distance of The portion of the denominator n sin ⁡ θ {\displaystyle n\sin \theta } is called the numerical aperture (NA) and can reach about 1.4–1.6 in modern optics, hence the Abbe limit is d = λ 2.8 {\displaystyle d={\frac {\lambda }{2.8}}} . The same formula had been proven by Hermann von Helmholtz in 1874. Considering green light around 500 nm and

SECTION 20

#1732858475246

2222-448: Is a topographic image. In other words, the topographic image is a trace of the tip of the probe regulated so that the df is constant and it may also be considered to be a plot of a constant-height surface of the df. Therefore, the topographic image of the AFM is not the exact surface morphology itself, but actually the image influenced by the bond-order between the probe and the sample, however,

2323-511: Is an earlier example of automatic control, but since it does not have an amplifier or gain , it is not usually considered a servomechanism. The first feedback position control device was the ship steering engine , used to position the rudder of large ships based on the position of the ship's wheel. John McFarlane Gray was a pioneer. His patented design was used on the SS Great Eastern in 1866. Joseph Farcot may deserve equal credit for

2424-410: Is combined with a rotary encoder or a potentiometer to form a servomechanism. This assembly may in turn form part of another servomechanism. A potentiometer provides a simple analog signal to indicate position, while an encoder provides position and usually speed feedback, which by the use of a PID controller allow more precise control of position and thus faster achievement of a stable position (for

2525-410: Is commonly achieved with a small piezo element in the cantilever holder, but other possibilities include an AC magnetic field (with magnetic cantilevers), piezoelectric cantilevers, or periodic heating with a modulated laser beam. The amplitude of this oscillation usually varies from several nm to 200 nm. In tapping mode, the frequency and amplitude of the driving signal are kept constant, leading to

2626-588: Is found by measuring the size of the beam at its waist, and its divergence far from the waist, and taking the product of the two, known as the beam parameter product . The ratio of this measured beam parameter product to that of the ideal is defined as M , so that M =1 describes an ideal beam. The M value of a beam is conserved when it is transformed by diffraction-limited optics. The outputs of many low and moderately powered lasers have M values of 1.2 or less, and are essentially diffraction-limited. The same equations apply to other wave-based sensors, such as radar and

2727-454: Is generally used in an open-loop manner without feedback. They are generally used for medium-precision applications. RC servos are used to provide actuation for various mechanical systems such as the steering of a car, the control surfaces on a plane, or the rudder of a boat. Due to their affordability, reliability, and simplicity of control by microprocessors, they are often used in small-scale robotics applications. A standard RC receiver (or

2828-403: Is gentle enough even for the visualization of supported lipid bilayers or adsorbed single polymer molecules (for instance, 0.4 nm thick chains of synthetic polyelectrolytes ) under liquid medium. With proper scanning parameters, the conformation of single molecules can remain unchanged for hours, and even single molecular motors can be imaged while moving. When operating in tapping mode,

2929-437: Is needed. By applying a small dither to the tip, the stiffness (force gradient) of the bond can be measured as well. Force spectroscopy is used in biophysics to measure the mechanical properties of living material (such as tissue or cells) or detect structures of different stiffness buried into the bulk of the sample using the stiffness tomography. Another application was to measure the interaction forces between from one hand

3030-431: Is only valid in the far field as it assumes that no evanescent fields reach the detector. Various near-field techniques that operate less than ≈1 wavelength of light away from the image plane can obtain substantially higher resolution. These techniques exploit the fact that the evanescent field contains information beyond the diffraction limit which can be used to construct very high resolution images, in principle beating

3131-452: Is possible over macroscopic distances. These techniques usually exploit optical nonlinearity in a material's reflected light to generate resolution beyond the diffraction limit. Among these techniques, the STED microscope has been one of the most successful. In STED, multiple laser beams are used to first excite, and then quench fluorescent dyes. The nonlinear response to illumination caused by

Atomic force microscopy - Misplaced Pages Continue

3232-581: Is proportional to the wavelength of the light being observed, and inversely proportional to the diameter of its objective 's entrance aperture . For telescopes with circular apertures, the size of the smallest feature in an image that is diffraction limited is the size of the Airy disk . As one decreases the size of the aperture of a telescopic lens , diffraction proportionately increases. At small apertures, such as f/22 , most modern lenses are limited only by diffraction and not by aberrations or other imperfections in

3333-481: Is rarely used. Further, under partially coherent conditions, the recorded image is often non-linear with object's scattering potential—especially when looking at non-self-luminous (non-fluorescent) objects. To boost contrast, and sometimes to linearize the system, unconventional microscopes (with structured illumination ) synthesize the condenser illumination by acquiring a sequence of images with known illumination parameters. Typically, these images are composited to form

3434-470: Is said to be diffraction-limited if it has reached this limit of resolution performance. Other factors may affect an optical system's performance, such as lens imperfections or aberrations , but these are caused by errors in the manufacture or calculation of a lens, whereas the diffraction limit is the maximum resolution possible for a theoretically perfect, or ideal, optical system. The diffraction-limited angular resolution , in radians, of an instrument

3535-780: Is similar to the pixel size for the majority of commercially available 'full frame' (43mm sensor diagonal) cameras and so these will operate in regime 3 for f-numbers around 8 (few lenses are close to diffraction limited at f-numbers smaller than 8). Cameras with smaller sensors will tend to have smaller pixels, but their lenses will be designed for use at smaller f-numbers and it is likely that they will also operate in regime 3 for those f-numbers for which their lenses are diffraction limited. There are techniques for producing images that appear to have higher resolution than allowed by simple use of diffraction-limited optics. Although these techniques improve some aspect of resolution, they generally come at an enormous increase in cost and complexity. Usually

3636-425: Is that laser beams are typically soft-edged beams. This non-uniformity in light distribution leads to a coefficient slightly different from the 1.22 value familiar in imaging. However, the scaling with wavelength and aperture is exactly the same. The beam quality of a laser beam is characterized by how well its propagation matches an ideal Gaussian beam at the same wavelength. The beam quality factor M squared (M )

3737-421: Is used as the parameter that goes into the electronic servo that controls the height of the cantilever above the sample. The servo adjusts the height to maintain a set cantilever oscillation amplitude as the cantilever is scanned over the sample. A tapping AFM image is therefore produced by imaging the force of the intermittent contacts of the tip with the sample surface. Although the peak forces applied during

3838-496: Is used to correct the action of the mechanism. In displacement-controlled applications, it usually includes a built-in encoder or other position feedback mechanism to ensure the output is achieving the desired effect. Following a specified motion trajectory is called servoing , where "servo" is used as a verb . The servo prefix originates from the Latin word servus meaning slave. The term correctly applies only to systems where

3939-432: Is vibrated or oscillated at a given frequency. AFM operation is usually described as one of three modes, according to the nature of the tip motion: contact mode, also called static mode (as opposed to the other two modes, which are called dynamic modes); tapping mode, also called intermittent contact, AC mode, or vibrating mode, or, after the detection mechanism, amplitude modulation AFM; and non-contact mode, or, again after

4040-506: The Earth work at a much lower resolution than the diffraction limit because of the distortion introduced by the passage of light through several kilometres of turbulent atmosphere. Advanced observatories have started using adaptive optics technology, resulting in greater image resolution for faint targets, but it is still difficult to reach the diffraction limit using adaptive optics. Radio telescopes are frequently diffraction-limited, because

4141-536: The feedback or error-correction signals help control mechanical position, speed, attitude or any other measurable variables. For example, an automotive power window control is not a servomechanism, as there is no automatic feedback that controls position—the operator does this by observation. By contrast a car's cruise control uses closed-loop feedback, which classifies it as a servomechanism. A common type of servo provides position control . Commonly, servos are electric , hydraulic , or pneumatic . They operate on

Atomic force microscopy - Misplaced Pages Continue

4242-409: The phase of oscillation can be used to discriminate between different types of materials on the surface. Amplitude modulation can be operated either in the non-contact or in the intermittent contact regime. In dynamic contact mode, the cantilever is oscillated such that the separation distance between the cantilever tip and the sample surface is modulated. Amplitude modulation has also been used in

4343-432: The AFM tip is extended towards and retracted from the surface as the deflection of the cantilever is monitored as a function of piezoelectric displacement. These measurements have been used to measure nanoscale contacts, atomic bonding , Van der Waals forces , and Casimir forces , dissolution forces in liquids and single molecule stretching and rupture forces. AFM has also been used to measure, in an aqueous environment,

4444-507: The Abbe diffraction limit formula. For instance, for an f/8 lens ( N = 8 {\displaystyle N=8} and N A ≈ 2.5 % {\displaystyle NA\approx 2.5\%} ) and for green light ( λ g = {\displaystyle \lambda _{g}=} 0.5 μm wavelength) light, the focusing spot diameter will be d = 9.76 μm or 19.5 λ g {\displaystyle \lambda _{g}} . This

4545-413: The aircraft's control surfaces, and radio-controlled models which use RC servos for the same purpose. Many autofocus cameras also use a servomechanism to accurately move the lens. A hard disk drive has a magnetic servo system with sub-micrometer positioning accuracy. In industrial machines, servos are used to perform complex motion, in many applications. A servomotor is a specific type of motor that

4646-445: The analyzes of the mean unfolding forces with the appropriate model leads to the obtainment of the information about the unfolding rate and free energy profile parameters of the protein. Diffraction limited In optics , any optical instrument or system  – a microscope , telescope , or camera  – has a principal limit to its resolution due to the physics of diffraction . An optical instrument

4747-440: The beam (by creating a vacuum) and staining the sample are not necessary. There are several types of scanning microscopy including SPM (which includes AFM, scanning tunneling microscopy (STM) and near-field scanning optical microscope (SNOM/NSOM), STED microscopy (STED), and scanning electron microscopy and electrochemical AFM , EC-AFM). Although SNOM and STED use visible , infrared or even terahertz light to illuminate

4848-399: The case of rigid samples, contact and non-contact images may look the same. However, if a few monolayers of adsorbed fluid are lying on the surface of a rigid sample, the images may look quite different. An AFM operating in contact mode will penetrate the liquid layer to image the underlying surface, whereas in non-contact mode an AFM will oscillate above the adsorbed fluid layer to image both

4949-414: The concavity and convexity accompanied with a scan of the sample along x–y direction (without height regulation in z-direction). As a result, the frequency shift arises. The image in which the values of the frequency obtained by a raster scan along the x–y direction of the sample surface are plotted against the x–y coordination of each measurement point is called a constant-height image. On the other hand,

5050-515: The construction. For microscopic instruments, the diffraction-limited spatial resolution is proportional to the light wavelength, and to the numerical aperture of either the objective or the object illumination source, whichever is smaller. In astronomy , a diffraction-limited observation is one that achieves the resolution of a theoretically ideal objective in the size of instrument used. However, most observations from Earth are seeing -limited due to atmospheric effects. Optical telescopes on

5151-413: The contacting part of the oscillation can be much higher than typically used in contact mode, tapping mode generally lessens the damage done to the surface and the tip compared to the amount done in contact mode. This can be explained by the short duration of the applied force, and because the lateral forces between tip and sample are significantly lower in tapping mode over contact mode. Tapping mode imaging

SECTION 50

#1732858475246

5252-505: The deflection (displacement with respect to the equilibrium position) of the cantilever and converts it into an electrical signal. The intensity of this signal will be proportional to the displacement of the cantilever. Various methods of detection can be used, e.g. interferometry, optical levers, the piezoelectric method, and STM-based detectors (see section "AFM cantilever deflection measurement"). This section applies specifically to imaging in § Contact mode . For other imaging modes,

5353-415: The deflection even when scanning in constant force mode, with feedback. This reveals the small tracking error of the feedback, and can sometimes reveal features that the feedback was not able to adjust for. The AFM signals, such as sample height or cantilever deflection, are recorded on a computer during the x–y scan. They are plotted in a pseudocolor image, in which each pixel represents an x–y position on

5454-403: The deflection of the cantilever is recorded as a function of the sample x–y position. As long as the tip is in contact with the sample, the deflection then corresponds to surface topography. This method is now less commonly used because the forces between tip and sample are not controlled, which can lead to forces high enough to damage the tip or the sample. It is, however, common practice to record

5555-408: The detection mechanism, frequency modulation AFM. Despite the nomenclature, repulsive contact can occur or be avoided both in amplitude modulation AFM and frequency modulation AFM, depending on the settings. In contact mode, the tip is "dragged" across the surface of the sample and the contours of the surface are measured either using the deflection of the cantilever directly or, more commonly, using

5656-420: The detector. The first one (using z-Feedback loop), said to be "constant XX mode" ( XX is something which kept by z-Feedback loop). Topographic image formation mode is based on abovementioned "constant XX mode", z-Feedback loop controls the relative distance between the probe and the sample through outputting control signals to keep constant one of frequency, vibration and phase which typically corresponds to

5757-417: The df may be kept constant by moving the probe upward and downward (See (3) of FIG.5) in z-direction using a negative feedback (by using z-feedback loop) while the raster scan of the sample surface along the x–y direction. The image in which the amounts of the negative feedback (the moving distance of the probe upward and downward in z-direction) are plotted against the x–y coordination of each measurement point

5858-529: The diffraction limit by a factor proportional to how well a specific imaging system can detect the near-field signal. For scattered light imaging, instruments such as near-field scanning optical microscopes and nano-FTIR , which are built atop atomic force microscope systems, can be used to achieve up to 10-50 nm resolution. The data recorded by such instruments often requires substantial processing, essentially solving an optical inverse problem for each image. Metamaterial -based superlenses can image with

5959-441: The diffraction-limited PSF is approximated by the diameter of the first null of the Airy disk , where λ {\displaystyle \lambda } is the wavelength of the light and N {\displaystyle N} is the f-number of the imaging optics, i.e., 2 N A → ( 2.44 N ) − 1 {\displaystyle 2NA\rightarrow (2.44N)^{-1}} in

6060-405: The dispersion force due to polymer adsorbed on the substrate. Forces of the order of a few piconewtons can now be routinely measured with a vertical distance resolution of better than 0.1 nanometers. Force spectroscopy can be performed with either static or dynamic modes. In dynamic modes, information about the cantilever vibration is monitored in addition to the static deflection. Problems with

6161-458: The distance along the z axis between the probe support (2 in fig. 3) and the sample support (8 in fig 3). As long as the tip remains in contact with the sample, and the sample is scanned in the x–y plane, height variations in the sample will change the deflection of the cantilever. The feedback then adjusts the height of the probe support so that the deflection is restored to a user-defined value (the setpoint). A properly adjusted feedback loop adjusts

SECTION 60

#1732858475246

6262-468: The drive attached to the sample, the drive can also be attached to the tip, or independent drives can be attached to both, since it is the relative displacement of the sample and tip that needs to be controlled. Controllers and plotter are not shown in Fig. 3. According to the configuration described above, the interaction between tip and sample, which can be an atomic-scale phenomenon, is transduced into changes of

6363-463: The error towards zero. The Ragonnet power reverse mechanism was a general purpose air or steam-powered servo amplifier for linear motion patented in 1909. Electrical servomechanisms were used as early as 1888 in Elisha Gray 's Telautograph . Electrical servomechanisms require a power amplifier. World War II saw the development of electrical fire-control servomechanisms, using an amplidyne as

6464-466: The expense of system complexity. Servomechanism In mechanical and control engineering , a servomechanism (also called servo system , or simply servo ) is a control system for the position and its time derivatives , such as velocity , of a mechanical system . It often includes a servomotor , and uses closed-loop control to reduce steady-state error and improve dynamic response. In closed-loop control, error-sensing negative feedback

6565-422: The feedback concept, with several patents between 1862 and 1868. The telemotor was invented around 1872 by Andrew Betts Brown , allowing elaborate mechanisms between the control room and the engine to be greatly simplified. Steam steering engines had the characteristics of a modern servomechanism: an input, an output, an error signal, and a means for amplifying the error signal used for negative feedback to drive

6666-399: The feedback signal required to keep the cantilever at a constant position. Because the measurement of a static signal is prone to noise and drift, low stiffness cantilevers (i.e. cantilevers with a low spring constant, k) are used to achieve a large enough deflection signal while keeping the interaction force low. Close to the surface of the sample, attractive forces can be quite strong, causing

6767-613: The field of solid state physics include (a) the identification of atoms at a surface, (b) the evaluation of interactions between a specific atom and its neighboring atoms, and (c) the study of changes in physical properties arising from changes in an atomic arrangement through atomic manipulation. In molecular biology, AFM can be used to study the structure and mechanical properties of protein complexes and assemblies. For example, AFM has been used to image microtubules and measure their stiffness. In cellular biology, AFM can be used to attempt to distinguish cancer cells and normal cells based on

6868-421: The frequency modulation mode allows for the use of very stiff cantilevers. Stiff cantilevers provide stability very close to the surface and, as a result, this technique was the first AFM technique to provide true atomic resolution in ultra-high vacuum conditions. In amplitude modulation, changes in the oscillation amplitude or phase provide the feedback signal for imaging. In amplitude modulation, changes in

6969-723: The human ear. As opposed to light waves (i.e., photons), massive particles have a different relationship between their quantum mechanical wavelength and their energy. This relationship indicates that the effective "de Broglie" wavelength is inversely proportional to the momentum of the particle. For example, an electron at an energy of 10 keV has a wavelength of 0.01 nm, allowing the electron microscope ( SEM or TEM ) to achieve high resolution images. Other massive particles such as helium, neon, and gallium ions have been used to produce images at resolutions beyond what can be attained with visible light. Such instruments provide nanometer scale imaging, analysis and fabrication capabilities at

7070-507: The instrument response function (IRF) can be approximated by a rectangle function, with a width equivalent to the pixel pitch. A more complete derivation of the modulation transfer function (derived from the PSF) of image sensors is given by Fliegel. Whatever the exact instrument response function, it is largely independent of the f-number of the lens. Thus at different f-numbers a camera may operate in three different regimes, as follows: The spread of

7171-402: The intensity of a value as a hue. Usually, the correspondence between the intensity of a value and a hue is shown as a color scale in the explanatory notes accompanying the image. Operation mode of image forming of the AFM are generally classified into two groups from the viewpoint whether it uses z-Feedback loop (not shown) to maintain the tip-sample distance to keep signal intensity exported by

7272-480: The liquid and surface. Schemes for dynamic mode operation include frequency modulation where a phase-locked loop is used to track the cantilever's resonance frequency and the more common amplitude modulation with a servo loop in place to keep the cantilever excitation to a defined amplitude. In frequency modulation, changes in the oscillation frequency provide information about tip-sample interactions. Frequency can be measured with very high sensitivity and thus

7373-405: The majority of SPM techniques are extensions of AFM that use this modality. The major difference between atomic force microscopy and competing technologies such as optical microscopy and electron microscopy is that AFM does not use lenses or beam irradiation. Therefore, it does not suffer from a limitation in spatial resolution due to diffraction and aberration, and preparing a space for guiding

7474-507: The motion of cantilever (for instance, voltage is applied to the Z-piezoelectric element and it moves the sample up and down towards the Z direction. When the distance between the probe and the sample is brought to the range where atomic force may be detected, while a cantilever is excited in its natural eigenfrequency ( f 0 ), the resonance frequency f of the cantilever may shift from its original resonance frequency. In other words, in

7575-399: The motion of cantilever, which is a macro-scale phenomenon. Several different aspects of the cantilever motion can be used to quantify the interaction between the tip and sample, most commonly the value of the deflection, the amplitude of an imposed oscillation of the cantilever, or the shift in resonance frequency of the cantilever (see section Imaging Modes). The detector (5) of AFM measures

7676-532: The name, the Atomic Force Microscope does not use the Nuclear force . The AFM has three major abilities: force measurement, topographic imaging, and manipulation. In force measurement, AFMs can be used to measure the forces between the probe and the sample as a function of their mutual separation. This can be applied to perform force spectroscopy , to measure the mechanical properties of the sample, such as

7777-469: The non-contact regime to image with atomic resolution by using very stiff cantilevers and small amplitudes in an ultra-high vacuum environment. Image formation is a plotting method that produces a color mapping through changing the x–y position of the tip while scanning and recording the measured variable, i.e. the intensity of control signal, to each x–y coordinate. The color mapping shows the measured value corresponding to each coordinate. The image expresses

7878-432: The phase of the cantilever's oscillation with respect to the driving signal can be recorded as well. This signal channel contains information about the energy dissipated by the cantilever in each oscillation cycle. Samples that contain regions of varying stiffness or with different adhesion properties can give a contrast in this channel that is not visible in the topographic image. Extracting the sample's material properties in

7979-604: The power amplifier. Vacuum tube amplifiers were used in the UNISERVO tape drive for the UNIVAC I computer. The Royal Navy began experimenting with Remote Power Control ( RPC ) on HMS Champion in 1928 and began using RPC to control searchlights in the early 1930s. During WW2 RPC was used to control gun mounts and gun directors. Modern servomechanisms use solid state power amplifiers, usually built from MOSFET or thyristor devices. Small servos may use power transistors . The origin of

8080-458: The principle of negative feedback, where the control input is compared to the actual position of the mechanical system as measured by some type of transducer at the output. Any difference between the actual and wanted values (an "error signal") is amplified (and converted) and used to drive the system in the direction necessary to reduce or eliminate the error. This procedure is one widely used application of control theory . Typical servos can give

8181-450: The process is similar, except that "deflection" should be replaced by the appropriate feedback variable. When using the AFM to image a sample, the tip is brought into contact with the sample, and the sample is raster scanned along an x–y grid (fig 4). Most commonly, an electronic feedback loop is employed to keep the probe-sample force constant during scanning. This feedback loop has the cantilever deflection as input, and its output controls

8282-419: The quenching process in which adding more light causes the image to become less bright generates sub-diffraction limited information about the location of dye molecules, allowing resolution far beyond the diffraction limit provided high illumination intensities are used. The limits on focusing or collimating a laser beam are very similar to the limits on imaging with a microscope or telescope. The only difference

8383-432: The range where atomic force may be detected, a frequency shift ( df  = f – f 0 ) will also be observed. When the distance between the probe and the sample is in the non-contact region, the frequency shift increases in negative direction as the distance between the probe and the sample gets smaller. When the sample has concavity and convexity, the distance between the tip-apex and the sample varies in accordance with

8484-402: The sample's Young's modulus , a measure of stiffness. For imaging, the reaction of the probe to the forces that the sample imposes on it can be used to form an image of the three-dimensional shape (topography) of a sample surface at a high resolution. This is achieved by raster scanning the position of the sample with respect to the tip and recording the height of the probe that corresponds to

8585-429: The sample, and the color represents the recorded signal. The AFM was invented by IBM scientists in 1985. The precursor to the AFM, the scanning tunneling microscope (STM), was developed by Gerd Binnig and Heinrich Rohrer in the early 1980s at IBM Research – Zurich , a development that earned them the 1986 Nobel Prize for Physics . Binnig invented the atomic force microscope and the first experimental implementation

8686-405: The sample, their resolution is not constrained by the diffraction limit. Fig. 3 shows an AFM, which typically consists of the following features. Numbers in parentheses correspond to numbered features in Fig. 3. Coordinate directions are defined by the coordinate system (0). The small spring-like cantilever (1) is carried by the support (2). Optionally, a piezoelectric element (typically made of

8787-404: The sample. In a digital camera, diffraction effects interact with the effects of the regular pixel grid. The combined effect of the different parts of an optical system is determined by the convolution of the point spread functions (PSF). The point spread function of a diffraction limited circular-aperture lens is simply the Airy disk . The point spread function of the camera, otherwise called

8888-463: The support-sample separation continuously during the scanning motion, such that the deflection remains approximately constant. In this situation, the feedback output equals the sample surface topography to within a small error. Historically, a different operation method has been used, in which the sample-probe support distance is kept constant and not controlled by a feedback ( servo mechanism ). In this mode, usually referred to as "constant-height mode",

8989-432: The surface presents a major problem for contact mode in ambient conditions. Dynamic contact mode (also called intermittent contact, AC mode or tapping mode) was developed to bypass this problem. Nowadays, tapping mode is the most frequently used AFM mode when operating in ambient conditions or in liquids. In tapping mode , the cantilever is driven to oscillate up and down at or near its resonance frequency. This oscillation

9090-419: The technique include no direct measurement of the tip-sample separation and the common need for low-stiffness cantilevers, which tend to "snap" to the surface. These problems are not insurmountable. An AFM that directly measures the tip-sample separation has been developed. The snap-in can be reduced by measuring in liquids or by using stiffer cantilevers, but in the latter case a more sensitive deflection sensor

9191-409: The technique is only appropriate for a small subset of imaging problems, with several general approaches outlined below. The effective resolution of a microscope can be improved by illuminating from the side. In conventional microscopes such as bright-field or differential interference contrast , this is achieved by using a condenser. Under spatially incoherent conditions, the image is understood as

9292-521: The tip is brought into proximity of a sample surface, forces between the tip and the sample lead to a deflection of the cantilever according to Hooke's law . Depending on the situation, forces that are measured in AFM include mechanical contact force, van der Waals forces , capillary forces , chemical bonding , electrostatic forces , magnetic forces (see magnetic force microscope , MFM), Casimir forces , solvation forces , etc. Along with force, additional quantities may simultaneously be measured through

9393-405: The tip to "snap-in" to the surface. Thus, contact mode AFM is almost always done at a depth where the overall force is repulsive, that is, in firm "contact" with the solid surface. In ambient conditions, most samples develop a liquid meniscus layer. Because of this, keeping the probe tip close enough to the sample for short-range forces to become detectable while preventing the tip from sticking to

9494-421: The tip to probe the electrical conductivity or transport of the underlying surface, but this is a challenging task with few research groups reporting consistent data (as of 2004). The AFM consists of a cantilever with a sharp tip (probe) at its end that is used to scan the specimen surface. The cantilever is typically silicon or silicon nitride with a tip radius of curvature on the order of nanometers. When

9595-419: The tip-to-sample distance at each (x,y) data point allows the scanning software to construct a topographic image of the sample surface. Non-contact mode AFM does not suffer from tip or sample degradation effects that are sometimes observed after taking numerous scans with contact AFM. This makes non-contact AFM preferable to contact AFM for measuring soft samples, e.g. biological samples and organic thin film. In

9696-412: The topographic image of the AFM is considered to reflect the geographical shape of the surface more than the topographic image of a scanning tunnel microscope. Besides imaging, AFM can be used for force spectroscopy , the direct measurement of tip-sample interaction forces as a function of the gap between the tip and sample. The result of this measurement is called a force-distance curve. For this method,

9797-401: The use of specialized types of probes (see scanning thermal microscopy , scanning joule expansion microscopy , photothermal microspectroscopy , etc.). The AFM can be operated in a number of modes, depending on the application. In general, possible imaging modes are divided into static (also called contact ) modes and a variety of dynamic (non-contact or "tapping") modes where the cantilever

9898-523: The wavelengths they use (from millimeters to meters) are so long that the atmospheric distortion is negligible. Space-based telescopes (such as Hubble , or a number of non-optical telescopes) always work at their diffraction limit, if their design is free of optical aberration . The beam from a laser with near-ideal beam propagation properties may be described as being diffraction-limited. A diffraction-limited laser beam, passed through diffraction-limited optics, will remain diffraction-limited, and will have

9999-538: The word is believed to come from the French " Le Servomoteur " or the slavemotor, first used by J. J. L. Farcot in 1868 to describe hydraulic and steam engines for use in ship steering. The simplest kind of servos use bang–bang control . More complex control systems use proportional control, PID control , and state space control, which are studied in modern control theory . Servos can be classified by means of their feedback control systems: The servo bandwidth indicates

10100-420: Was demonstrated in 1993 by Ohnesorge and Binnig. True atomic resolution of the silicon 7x7 surface—the atomic images of this surface obtained by STM had convinced the scientific community of the spectacular spatial resolution of scanning tunneling microscopy—had to wait a little longer before it was shown by Giessibl. Subatomic resolution (i.e. the ability to resolve structural details within the electron density of

10201-557: Was made by Binnig, Quate and Gerber in 1986. The first commercially available atomic force microscope was introduced in 1989. The AFM is one of the foremost tools for imaging, measuring, and manipulating matter at the nanoscale . The AFM has been applied to problems in a wide range of disciplines of the natural sciences, including solid-state physics , semiconductor science and technology, molecular engineering , polymer chemistry and physics , surface chemistry , molecular biology , cell biology , and medicine . Applications in

#245754