Misplaced Pages

Fourier-transform ion cyclotron resonance

Article snapshot taken from Wikipedia with creative commons attribution-sharealike license. Give it a read and then ask your questions in the chat. We can research this topic together.

Fourier-transform ion cyclotron resonance mass spectrometry is a type of mass analyzer (or mass spectrometer ) for determining the mass-to-charge ratio ( m / z ) of ions based on the cyclotron frequency of the ions in a fixed magnetic field. The ions are trapped in a Penning trap (a magnetic field with electric trapping plates), where they are excited (at their resonant cyclotron frequencies) to a larger cyclotron radius by an oscillating electric field orthogonal to the magnetic field. After the excitation field is removed, the ions are rotating at their cyclotron frequency in phase (as a "packet" of ions). These ions induce a charge (detected as an image current) on a pair of electrodes as the packets of ions pass close to them. The resulting signal is called a free induction decay (FID), transient or interferogram that consists of a superposition of sine waves . The useful signal is extracted from this data by performing a Fourier transform to give a mass spectrum .

#752247

81-559: FT-ICR was invented by Melvin B. Comisarow and Alan G. Marshall at the University of British Columbia . The first paper appeared in Chemical Physics Letters in 1974. The inspiration was earlier developments in conventional ICR and Fourier-transform nuclear magnetic resonance (FT-NMR) spectrometry. Marshall has continued to develop the technique at The Ohio State University and Florida State University . The physics of FTICR

162-486: A mass spectrum , a record of ions as a function of m/Q . Typically, some type of electron multiplier is used, though other detectors including Faraday cups and ion-to-photon detectors are also used. Because the number of ions leaving the mass analyzer at a particular instant is typically quite small, considerable amplification is often necessary to get a signal. Microchannel plate detectors are commonly used in modern commercial instruments. In FTMS and Orbitraps ,

243-470: A spectrum of mass values on a photographic plate . A mass spectroscope is similar to a mass spectrograph except that the beam of ions is directed onto a phosphor screen. A mass spectroscope configuration was used in early instruments when it was desired that the effects of adjustments be quickly observed. Once the instrument was properly adjusted, a photographic plate was inserted and exposed. The term mass spectroscope continued to be used even though

324-400: A broad application, in practice have come instead to connote a specific or a limited number of instrument configurations. An example of this is isotope-ratio mass spectrometry (IRMS), which refers in practice to the use of a limited number of sector based mass analyzers; this name is used to refer to both the application and the instrument used for the application. An important enhancement to

405-443: A central, spindle shaped electrode. The electrode confines the ions so that they both orbit around the central electrode and oscillate back and forth along the central electrode's long axis. This oscillation generates an image current in the detector plates which is recorded by the instrument. The frequencies of these image currents depend on the mass-to-charge ratios of the ions. Mass spectra are obtained by Fourier transformation of

486-721: A distribution of isotopes that produce a series of isotopic peaks. Because the isotopic peaks are close to each other on the m / z axis, due to the multiple charges, the high resolving power of the FTICR is extremely useful. FTICR-MS is very useful in other studies of proteomics as well. It achieves exceptional resolution in both top-down and bottom-up proteomics. Electron-capture dissociation (ECD), collisional-induced dissociation (CID), and infrared multiphoton dissociation (IRMPD) are all utilized to produce fragment spectra in tandem mass spectrometry experiments. Although CID and IRMPD use vibrational excitation to further dissociate peptides by breaking

567-420: A large discrimination in kinetic ion acceleration between positive and negative ions trapped simultaneously inside the new cell. Several ion axial acceleration schemes were recently written for ion–ion collision studies. Stored-waveform inverse Fourier transform (SWIFT) is a method for the creation of excitation waveforms for FTMS. The time-domain excitation waveform is formed from the inverse Fourier transform of

648-418: A magnetic field. Thus, the different ions are not detected in different places as with sector instruments or at different times as with time-of-flight instruments, but all ions are detected simultaneously during the detection interval. This provides an increase in the observed signal-to-noise ratio owing to the principles of Fellgett's advantage . In FTICR-MS, resolution can be improved either by increasing

729-414: A mass filter, to transmit a particular fragment ion to the detector. If a quadrupole is made to rapidly and repetitively cycle through a range of mass filter settings, full spectra can be reported. Likewise, a triple quad can be made to perform various scan types characteristic of tandem mass spectrometry . The quadrupole ion trap works on the same physical principles as the quadrupole mass analyzer, but

810-775: A mass spectrometer. A collision cell then stabilizes the peptide ions while they collide with a gas, causing them to fragment by collision-induced dissociation (CID). A further mass analyzer then sorts the fragments produced from the peptides. Tandem MS can also be done in a single mass analyzer over time, as in a quadrupole ion trap . There are various methods for fragmenting molecules for tandem MS, including collision-induced dissociation (CID), electron capture dissociation (ECD), electron transfer dissociation (ETD), infrared multiphoton dissociation (IRMPD), blackbody infrared radiative dissociation (BIRD), electron-detachment dissociation (EDD) and surface-induced dissociation (SID). An important application using tandem mass spectrometry

891-471: A natural abundance of about 25 percent). The analyzer part of the spectrometer contains electric and magnetic fields, which exert forces on ions traveling through these fields. The speed of a charged particle may be increased or decreased while passing through the electric field, and its direction may be altered by the magnetic field. The magnitude of the deflection of the moving ion's trajectory depends on its mass-to-charge ratio. Lighter ions are deflected by

SECTION 10

#1732855374753

972-418: A plot of intensity as a function of the mass-to-charge ratio. Mass spectrometry is used in many different fields and is applied to pure samples as well as complex mixtures. A mass spectrum is a type of plot of the ion signal as a function of the mass-to-charge ratio. These spectra are used to determine the elemental or isotopic signature of a sample, the masses of particles and of molecules , and to elucidate

1053-435: A range of m/z to catalog the ions present. The time-of-flight (TOF) analyzer uses an electric field to accelerate the ions through the same potential , and then measures the time they take to reach the detector. If the particles all have the same charge , their kinetic energies will be identical, and their velocities will depend only on their masses . For example, ions with a lower mass will travel faster, reaching

1134-413: A static electric and/or magnetic field to affect the path and/or velocity of the charged particles in some way. As shown above, sector instruments bend the trajectories of the ions as they pass through the mass analyzer, according to their mass-to-charge ratios, deflecting the more charged and faster-moving, lighter ions more. The analyzer can be used to select a narrow range of m/z or to scan through

1215-460: A wide array of sample types. In this source, a plasma that is electrically neutral overall, but that has had a substantial fraction of its atoms ionized by high temperature, is used to atomize introduced sample molecules and to further strip the outer electrons from those atoms. The plasma is usually generated from argon gas, since the first ionization energy of argon atoms is higher than the first of any other elements except He, F and Ne, but lower than

1296-475: Is MALDI-TOF , which refers to a combination of a matrix-assisted laser desorption/ionization source with a time-of-flight mass analyzer. Other examples include inductively coupled plasma-mass spectrometry (ICP-MS) , accelerator mass spectrometry (AMS) , thermal ionization-mass spectrometry (TIMS) and spark source mass spectrometry (SSMS) . Certain applications of mass spectrometry have developed monikers that although strictly speaking would seem to refer to

1377-540: Is a Canadian physicist and analytical chemist who co-invented the Fourier-transform ion cyclotron resonance technique of Mass spectroscopy , together with Alan G. Marshall , at the University of British Columbia . Comisarow was born in Alberta to a Ukrainian-Canadian family, and earned his bachelor's degree at the University of Alberta , 1963, before obtaining his PhD at Case Western Reserve University , under

1458-443: Is a wide variety of ionization techniques, depending on the phase (solid, liquid, gas) of the sample and the efficiency of various ionization mechanisms for the unknown species. An extraction system removes ions from the sample, which are then targeted through the mass analyzer and into the detector . The differences in masses of the fragments allows the mass analyzer to sort the ions by their mass-to-charge ratio. The detector measures

1539-492: Is an example of the linear ion trap. A toroidal ion trap can be visualized as a linear quadrupole curved around and connected at the ends or as a cross-section of a 3D ion trap rotated on edge to form the toroid, donut-shaped trap. The trap can store large volumes of ions by distributing them throughout the ring-like trap structure. This toroidal shaped trap is a configuration that allows the increased miniaturization of an ion trap mass analyzer. Additionally, all ions are stored in

1620-448: Is applied to the endcap electrodes, and the trapping voltage amplitude and/or excitation voltage frequency is varied to bring ions into a resonance condition in order of their mass/charge ratio. The cylindrical ion trap mass spectrometer (CIT) is a derivative of the quadrupole ion trap where the electrodes are formed from flat rings rather than hyperbolic shaped electrodes. The architecture lends itself well to miniaturization because as

1701-447: Is applied. This filament emits electrons which ionize the compounds. The ions can then further fragment, yielding predictable patterns. Intact ions and fragments pass into the mass spectrometer's analyzer and are eventually detected. However, the high temperatures (300 °C) used in the GC-MS injection port (and oven) can result in thermal degradation of injected molecules, thus resulting in

SECTION 20

#1732855374753

1782-457: Is commonly used for applying radial excitation electric field and detection. DC electric voltage is applied on the terminal ring electrodes to trap ions along the magnetic field lines. Open cylindrical cells with ring electrodes of different diameters have also been designed. They proved not only capable in trapping and detecting both ion polarities simultaneously, but also they succeeded to separate positive from negative ions radially. This presented

1863-407: Is designed to pass the untrapped ions rather than collect the trapped ones, and is for that reason referred to as a transmission quadrupole. A magnetically enhanced quadrupole mass analyzer includes the addition of a magnetic field, either applied axially or transversely. This novel type of instrument leads to an additional performance enhancement in terms of resolution and/or sensitivity depending upon

1944-474: Is in protein identification. Tandem mass spectrometry enables a variety of experimental sequences. Many commercial mass spectrometers are designed to expedite the execution of such routine sequences as selected reaction monitoring (SRM), precursor ion scanning, product ion scanning, and neutral loss scanning. Another type of tandem mass spectrometry used for radiocarbon dating is accelerator mass spectrometry (AMS), which uses very high voltages, usually in

2025-408: Is not suitable for coupling to HPLC , i.e. LC-MS , since at atmospheric pressure, the filaments used to generate electrons burn out rapidly. Thus EI is coupled predominantly with GC , i.e. GC-MS , where the entire system is under high vacuum. Hard ionization techniques are processes which impart high quantities of residual energy in the subject molecule invoking large degrees of fragmentation (i.e.

2106-442: Is similar to that of a cyclotron at least in the first approximation. In the simplest idealized form, the relationship between the cyclotron frequency and the mass-to-charge ratio is given by where f = cyclotron frequency, q = ion charge, B = magnetic field strength and m = ion mass. This is more often represented in angular frequency : where ω c {\displaystyle \omega _{\text{c}}}

2187-539: Is small, which is generally true. In that case the value of the radical is just slightly less than ω c / 2 {\displaystyle \omega _{\text{c}}/2} , and the value of ω + {\displaystyle \omega _{+}} is just slightly less than ω c {\displaystyle \omega _{\text{c}}} (the cyclotron frequency has been slightly reduced). For ω − {\displaystyle \omega _{-}}

2268-403: Is the angular cyclotron frequency , which is related to frequency by the definition f = ω 2 π {\displaystyle f={\frac {\omega }{2\pi }}} . Because of the quadrupolar electrical field used to trap the ions in the axial direction, this relationship is only approximate. The axial electrical trapping results in axial oscillations within

2349-485: Is the ratio of the m/z measurement error to the true m/z . Mass accuracy is usually measured in ppm or milli mass units . The mass range is the range of m/z amenable to analysis by a given analyzer. The linear dynamic range is the range over which ion signal is linear with analyte concentration. Speed refers to the time frame of the experiment and ultimately is used to determine the number of spectra per unit time that can be generated. A sector field mass analyzer uses

2430-463: Is the reduced cyclotron (angular) frequency and ω − {\displaystyle \omega _{-}} is the magnetron (angular) frequency. Again, ω + {\displaystyle \omega _{+}} is what is typically measured in FTICR. The meaning of this equation can be understood qualitatively by considering the case where ω t {\displaystyle \omega _{\text{t}}}

2511-636: Is used to dissociate stable gaseous molecules in a carrier gas of He or Ar. In instances where a synchrotron light source is utilized, a tuneable photon energy can be utilized to acquire a photoionization efficiency curve which can be used in conjunction with the charge ratio m/z to fingerprint molecular and ionic species. More recently atmospheric pressure photoionization (APPI) has been developed to ionize molecules mostly as effluents of LC-MS systems. Some applications for ambient ionization include environmental applications as well as clinical applications. In these techniques, ions form in an ion source outside

Fourier-transform ion cyclotron resonance - Misplaced Pages Continue

2592-673: The isotopes of uranium during the Manhattan Project . Calutron mass spectrometers were used for uranium enrichment at the Oak Ridge, Tennessee Y-12 plant established during World War II. In 1989, half of the Nobel Prize in Physics was awarded to Hans Dehmelt and Wolfgang Paul for the development of the ion trap technique in the 1950s and 1960s. In 2002, the Nobel Prize in Chemistry

2673-421: The (officially) dimensionless m/z , where z is the number of elementary charges ( e ) on the ion (z=Q/e). This quantity, although it is informally called the mass-to-charge ratio, more accurately speaking represents the ratio of the mass number and the charge number, z . There are many types of mass analyzers, using either static or dynamic fields, and magnetic or electric fields, but all operate according to

2754-656: The Spectroscopy Society of Canada (1989); 1995 Field Franklin Award for Mass Spectroscopy, from the American Chemical Society; and the 1996 Fisher Award in Analytical Chemistry of Canadian Society for Chemistry. Mass spectrometry Mass spectrometry ( MS ) is an analytical technique that is used to measure the mass-to-charge ratio of ions . The results are presented as a mass spectrum ,

2835-475: The above differential equation. Each analyzer type has its strengths and weaknesses. Many mass spectrometers use two or more mass analyzers for tandem mass spectrometry (MS/MS) . In addition to the more common mass analyzers listed below, there are others designed for special situations. There are several important analyzer characteristics. The mass resolving power is the measure of the ability to distinguish two peaks of slightly different m/z . The mass accuracy

2916-406: The above expressions for the force applied to the ion yields: This differential equation is the classic equation of motion for charged particles . Together with the particle's initial conditions, it completely determines the particle's motion in space and time in terms of m/Q . Thus mass spectrometers could be thought of as "mass-to-charge spectrometers". When presenting data, it is common to use

2997-451: The amount that the cyclotron frequency was reduced by). FTICR-MS differs significantly from other mass spectrometry techniques in that the ions are not detected by hitting a detector such as an electron multiplier but only by passing near detection plates. Additionally the masses are not resolved in space or time as with other techniques but only by the ion cyclotron resonance (rotational) frequency that each ion produces as it rotates in

3078-494: The analyte is ionized by chemical ion-molecule reactions during collisions in the source. Two techniques often used with liquid and solid biological samples include electrospray ionization (invented by John Fenn ) and matrix-assisted laser desorption/ionization (MALDI, initially developed as a similar technique "Soft Laser Desorption (SLD)" by K. Tanaka for which a Nobel Prize was awarded and as MALDI by M. Karas and F. Hillenkamp ). In mass spectrometry, ionization refers to

3159-443: The appropriate frequency-domain excitation spectrum, which is chosen to excite the resonance frequencies of selected ions. The SWIFT procedure can be used to select ions for tandem mass spectrometry experiments. Fourier-transform ion cyclotron resonance (FTICR) mass spectrometry is a high-resolution technique that can be used to determine masses with high accuracy. Many applications of FTICR-MS use this mass accuracy to help determine

3240-486: The backbone amide linkages, which are typically low in energy and weak, CID and IRMPD may also cause dissociation of post-translational modifications. ECD, on the other hand, allows specific modifications to be preserved. This is quite useful in analyzing phosphorylation states, O- or N-linked glycosylation, and sulfating. Melvin Barnett Comisarow Melvin Barnett Comisarow FRSC

3321-415: The central location of the peaks, since the starting velocity of ions is generally centered at zero. To fix this problem, time-lag focusing/ delayed extraction has been coupled with TOF-MS. Quadrupole mass analyzers use oscillating electrical fields to selectively stabilize or destabilize the paths of ions passing through a radio frequency (RF) quadrupole field created between four parallel rods. Only

Fourier-transform ion cyclotron resonance - Misplaced Pages Continue

3402-556: The chemical identity or structure of molecules and other chemical compounds . In a typical MS procedure, a sample, which may be solid, liquid, or gaseous, is ionized , for example by bombarding it with a beam of electrons . This may cause some of the sample's molecules to break up into positively charged fragments or simply become positively charged without fragmenting. These ions (fragments) are then separated according to their mass-to-charge ratio, for example by accelerating them and subjecting them to an electric or magnetic field: ions of

3483-441: The composition of molecules based on accurate mass. This is possible due to the mass defect of the elements. FTICR-MS is able to achieve higher levels of mass accuracy than other forms of mass spectrometer , in part, because a superconducting magnet is much more stable than radio-frequency (RF) voltage. Another place that FTICR-MS is useful is in dealing with complex mixtures, such as biomass or waste liquefaction products, since

3564-517: The detector consists of a pair of metal surfaces within the mass analyzer/ion trap region which the ions only pass near as they oscillate. No direct current is produced, only a weak AC image current is produced in a circuit between the electrodes. Other inductive detectors have also been used. A tandem mass spectrometer is one capable of multiple rounds of mass spectrometry, usually separated by some form of molecule fragmentation. For example, one mass analyzer can isolate one peptide from many entering

3645-439: The detector first. Ions usually are moving prior to being accelerated by the electric field , this causes particles with the same m/z to arrive at different times at the detector. This difference in initial velocities is often not dependent on the mass of the ion, and will turn into a difference in the final velocity. This distribution in velocities broadens the peaks shown on the count vs m/z plot, but will generally not change

3726-424: The detector is located. Ions of different mass are resolved according to impact time. The final element of the mass spectrometer is the detector. The detector records either the charge induced or the current produced when an ion passes by or hits a surface. In a scanning instrument, the signal produced in the detector during the course of the scan versus where the instrument is in the scan (at what m/Q ) will produce

3807-550: The direct illumination of a phosphor screen was replaced by indirect measurements with an oscilloscope . The use of the term mass spectroscopy is now discouraged due to the possibility of confusion with light spectroscopy . Mass spectrometry is often abbreviated as mass-spec or simply as MS . Modern techniques of mass spectrometry were devised by Arthur Jeffrey Dempster and F.W. Aston in 1918 and 1919 respectively. Sector mass spectrometers known as calutrons were developed by Ernest O. Lawrence and used for separating

3888-400: The discharge tube. English scientist J. J. Thomson later improved on the work of Wien by reducing the pressure to create the mass spectrograph. The word spectrograph had become part of the international scientific vocabulary by 1884. Early spectrometry devices that measured the mass-to-charge ratio of ions were called mass spectrographs which consisted of instruments that recorded

3969-459: The identification of known molecules it is also useful for identifying unknowns using its similarity searching/analysis. All tandem mass spectrometry data comes from the experimental analysis of standards at multiple collision energies and in both positive and negative ionization modes. When a specific combination of source, analyzer, and detector becomes conventional in practice, a compound acronym may arise to designate it succinctly. One example

4050-411: The identified masses or through a characteristic fragmentation pattern. In 1886, Eugen Goldstein observed rays in gas discharges under low pressure that traveled away from the anode and through channels in a perforated cathode , opposite to the direction of negatively charged cathode rays (which travel from cathode to anode). Goldstein called these positively charged anode rays "Kanalstrahlen";

4131-403: The ions according to their mass-to-charge ratio . The following two laws govern the dynamics of charged particles in electric and magnetic fields in vacuum: Here F is the force applied to the ion, m is the mass of the ion, a is the acceleration, Q is the ion charge, E is the electric field, and v × B is the vector cross product of the ion velocity and the magnetic field Equating

SECTION 50

#1732855374753

4212-416: The ions are injected into a Penning trap (a static electric/magnetic ion trap ) where they effectively form part of a circuit. Detectors at fixed positions in space measure the electrical signal of ions which pass near them over time, producing a periodic signal. Since the frequency of an ion's cycling is determined by its mass-to-charge ratio, this can be deconvoluted by performing a Fourier transform on

4293-559: The ions are trapped and sequentially ejected. Ions are trapped in a mainly quadrupole RF field, in a space defined by a ring electrode (usually connected to the main RF potential) between two endcap electrodes (typically connected to DC or auxiliary AC potentials). The sample is ionized either internally (e.g. with an electron or laser beam), or externally, in which case the ions are often introduced through an aperture in an endcap electrode. There are many mass/charge separation and isolation methods but

4374-417: The ions in a certain range of mass/charge ratio are passed through the system at any time, but changes to the potentials on the rods allow a wide range of m/z values to be swept rapidly, either continuously or in a succession of discrete hops. A quadrupole mass analyzer acts as a mass-selective filter and is closely related to the quadrupole ion trap , particularly the linear quadrupole ion trap except that it

4455-502: The isotopic composition of its constituents (the ratio of Cl to Cl). The ion source is the part of the mass spectrometer that ionizes the material under analysis (the analyte). The ions are then transported by magnetic or electric fields to the mass analyzer. Techniques for ionization have been key to determining what types of samples can be analyzed by mass spectrometry. Electron ionization and chemical ionization are used for gases and vapors . In chemical ionization sources,

4536-412: The magnetic field lines). Ions can be either generated inside the cell or can be injected to the cell from an external ionization source . Nested ICR cells with double pair of grids were also fabricated to trap both positive and negative ions simultaneously. The most common open cell geometry is a cylinder, which is axially segmented to produce electrodes in the shape of a ring. The central ring electrode

4617-408: The magnetic force to a greater degree than heavier ions (based on Newton's second law of motion , F = ma ). The streams of magnetically sorted ions pass from the analyzer to the detector, which records the relative abundance of each ion type. This information is used to determine the chemical element composition of the original sample (i.e. that both sodium and chlorine are present in the sample) and

4698-421: The magnetron frequency. The cyclotron motion is still the frequency being used, but the relationship above is not exact due to this phenomenon. The natural angular frequencies of motion are where ω t {\displaystyle \omega _{\text{t}}} is the axial trapping frequency due the axial electrical trapping and ω + {\displaystyle \omega _{+}}

4779-415: The magnitude and orientation of the applied magnetic field. A common variation of the transmission quadrupole is the triple quadrupole mass spectrometer. The "triple quad" has three consecutive quadrupole stages, the first acting as a mass filter to transmit a particular incoming ion to the second quadrupole, a collision chamber, wherein that ion can be broken into fragments. The third quadrupole also acts as

4860-430: The mass resolving and mass determining capabilities of mass spectrometry is using it in tandem with chromatographic and other separation techniques. A common combination is gas chromatography-mass spectrometry (GC/MS or GC-MS). In this technique, a gas chromatograph is used to separate different compounds. This stream of separated compounds is fed online into the ion source, a metallic filament to which voltage

4941-840: The mass spectrometer. Sampling becomes easy as the samples don't need previous separation nor preparation. Some examples of ambient ionization techniques are Direct Analysis in Real Time (DART), DESI , SESI , LAESI , desorption atmospheric-pressure chemical ionization (DAPCI), Soft Ionization by Chemical Reaction in Transfer (SICRT) and desorption atmospheric pressure photoionization DAPPI among others. Others include glow discharge , field desorption (FD), fast atom bombardment (FAB), thermospray , desorption/ionization on silicon (DIOS), atmospheric pressure chemical ionization (APCI), secondary ion mass spectrometry (SIMS), spark ionization and thermal ionization (TIMS). Mass analyzers separate

SECTION 60

#1732855374753

5022-467: The mega-volt range, to accelerate negative ions into a type of tandem mass spectrometer. The METLIN Metabolite and Chemical Entity Database is the largest repository of experimental tandem mass spectrometry data acquired from standards. The tandem mass spectrometry data on over 930,000 molecular standards (as of January 2024) is provided to facilitate the identification of chemical entities from tandem mass spectrometry experiments. In addition to

5103-407: The most commonly used is the mass instability mode in which the RF potential is ramped so that the orbit of ions with a mass a > b are stable while ions with mass b become unstable and are ejected on the z -axis onto a detector. There are also non-destructive analysis methods. Ions may also be ejected by the resonance excitation method, whereby a supplemental oscillatory excitation voltage

5184-617: The production of gas phase ions suitable for resolution in the mass analyser or mass filter. Ionization occurs in the ion source . There are several ion sources available; each has advantages and disadvantages for particular applications. For example, electron ionization (EI) gives a high degree of fragmentation, yielding highly detailed mass spectra which when skilfully analysed can provide important information for structural elucidation/characterisation and facilitate identification of unknown compounds by comparison to mass spectral libraries obtained under identical operating conditions. However, EI

5265-435: The recorded image currents. Orbitraps have a high mass accuracy, high sensitivity and a good dynamic range. Fourier-transform mass spectrometry (FTMS), or more precisely Fourier-transform ion cyclotron resonance MS, measures mass by detecting the image current produced by ions cyclotroning in the presence of a magnetic field. Instead of measuring the deflection of ions with a detector such as an electron multiplier ,

5346-411: The resolution (narrow peak width) allows the signals of two ions with similar mass-to-charge ratios ( m / z ) to be detected as distinct ions. This high resolution is also useful in studying large macromolecules such as proteins with multiple charges, which can be produced by electrospray ionization . For example, attomole level of detection of two peptides has been reported. These large molecules contain

5427-416: The same mass-to-charge ratio will undergo the same amount of deflection. The ions are detected by a mechanism capable of detecting charged particles, such as an electron multiplier . Results are displayed as spectra of the signal intensity of detected ions as a function of the mass-to-charge ratio. The atoms or molecules in the sample can be identified by correlating known masses (e.g. an entire molecule) to

5508-633: The same trapping field and ejected together simplifying detection that can be complicated with array configurations due to variations in detector alignment and machining of the arrays. As with the toroidal trap, linear traps and 3D quadrupole ion traps are the most commonly miniaturized mass analyzers due to their high sensitivity, tolerance for mTorr pressure, and capabilities for single analyzer tandem mass spectrometry (e.g. product ion scans). Orbitrap instruments are similar to Fourier-transform ion cyclotron resonance mass spectrometers (see text below). Ions are electrostatically trapped in an orbit around

5589-423: The sample is vaporized (turned into gas ) and ionized (transformed into electrically charged particles) into sodium (Na ) and chloride (Cl ) ions. Sodium atoms and ions are monoisotopic , with a mass of about 23 daltons (symbol: Da or older symbol: u). Chloride atoms and ions come in two stable isotopes with masses of approximately 35 u (at a natural abundance of about 75 percent) and approximately 37 u (at

5670-403: The second ionization energy of all except the most electropositive metals. The heating is achieved by a radio-frequency current passed through a coil surrounding the plasma. Photoionization can be used in experiments which seek to use mass spectrometry as a means of resolving chemical kinetics mechanisms and isomeric product branching. In such instances a high energy photon, either X-ray or uv,

5751-410: The signal. FTMS has the advantage of high sensitivity (since each ion is "counted" more than once) and much higher resolution and thus precision. Ion cyclotron resonance (ICR) is an older mass analysis technique similar to FTMS except that ions are detected with a traditional detector. Ions trapped in a Penning trap are excited by an RF electric field until they impact the wall of the trap, where

5832-436: The size of a trap is reduced, the shape of the electric field near the center of the trap, the region where the ions are trapped, forms a shape similar to that of a hyperbolic trap. A linear quadrupole ion trap is similar to a quadrupole ion trap, but it traps ions in a two dimensional quadrupole field, instead of a three-dimensional quadrupole field as in a 3D quadrupole ion trap. Thermo Fisher's LTQ ("linear trap quadrupole")

5913-402: The standard translation of this term into English is " canal rays ". Wilhelm Wien found that strong electric or magnetic fields deflected the canal rays and, in 1899, constructed a device with perpendicular electric and magnetic fields that separated the positive rays according to their charge-to-mass ratio ( Q/m ). Wien found that the charge-to-mass ratio depended on the nature of the gas in

5994-511: The strength of the magnet (in teslas ) or by increasing the detection duration. A review of different cell geometries with their specific electric configurations is available in the literature. However, ICR cells can belong to one of the following two categories: closed cells or open cells. Several closed ICR cells with different geometries were fabricated and their performance has been characterized. Grids were used as end caps to apply an axial electric field for trapping ions axially (parallel to

6075-462: The subject molecule and as such result in little fragmentation. Examples include fast atom bombardment (FAB), chemical ionization (CI), atmospheric-pressure chemical ionization (APCI), atmospheric-pressure photoionization (APPI), electrospray ionization (ESI), desorption electrospray ionization (DESI), and matrix-assisted laser desorption/ionization (MALDI). Inductively coupled plasma (ICP) sources are used primarily for cation analysis of

6156-643: The supervision of George Andrew Olah in 1969, and subsequently a postdoc with John D. Baldeschwieler at Stanford University . His first academic appointment was at the University of British Columbia, where he subsequently stayed until retirement. He is a fellow of the American Chemical Society , and the Royal Society of Canada , and has received numerous awards, including the Barringer Award of

6237-405: The systematic rupturing of bonds acts to remove the excess energy, restoring stability to the resulting ion). Resultant ions tend to have m/z lower than the molecular ion (other than in the case of proton transfer and not including isotope peaks). The most common example of hard ionization is electron ionization (EI). Soft ionization refers to the processes which impart little residual energy onto

6318-419: The trap with the (angular) frequency where α {\displaystyle \alpha } is a constant similar to the spring constant of a harmonic oscillator and is dependent on applied voltage, trap dimensions and trap geometry. The electric field and the resulting axial harmonic motion reduces the cyclotron frequency and introduces a second radial motion called magnetron motion that occurs at

6399-401: The value of an indicator quantity and thus provides data for calculating the abundances of each ion present. Some detectors also give spatial information, e.g., a multichannel plate. The following describes the operation of a spectrometer mass analyzer, which is of the sector type. (Other analyzer types are treated below.) Consider a sample of sodium chloride (table salt). In the ion source,

6480-469: The value of the radical is the same (slightly less than ω c / 2 {\displaystyle \omega _{\text{c}}/2} ), but it is being subtracted from ω c / 2 {\displaystyle \omega _{\text{c}}/2} , resulting in a small number equal to ω c − ω + {\displaystyle \omega _{\text{c}}-\omega _{+}} (i.e.

6561-425: Was awarded to John Bennett Fenn for the development of electrospray ionization (ESI) and Koichi Tanaka for the development of soft laser desorption (SLD) and their application to the ionization of biological macromolecules , especially proteins . A mass spectrometer consists of three components: an ion source, a mass analyzer, and a detector. The ionizer converts a portion of the sample into ions. There

#752247